How to Wear a Surgical Mask

Instructions for Use — USM1001 3-Ply Surgical Mask

A surgical mask only works the way it is supposed to when it is put on, worn and taken off correctly. These instructions apply to the USA MedPro

USM1001 3-ply earloop surgical mask, manufactured in Houston, Texas.

Read these instructions before use. Keep them with the carton, or link to this page from your own procedures.

What this mask is for

The USM1001 is a single-use, non-sterile surgical face mask. It is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material, and is intended for use in infection control practices to reduce the potential exposure to blood and body fluids.

The mask is supplied over-the-counter. It is worn by healthcare personnel and by members of the general public, and it is not restricted to aparticular use environment.

The mask works as a passive physical barrier. The outer and inner layers resist liquid; the middle layer filters particles and microorganisms. Worn over the nose and mouth, it limits the transfer of microorganisms, body fluids and particulate material between the wearer and the surrounding environment — including, during a surgical or other clinical procedure, to the surgical field and to other people present. A formable nose wire along the top edge lets you shape the mask to the bridge of your nose.

This mask is not a respirator. It does not provide the seal or the respiratory protection of a NIOSH-approved N95 or other filtering facepiecerespirator, and it should not be substituted for one where respiratory protection is required.

Before you begin

  • Check that the carton and the inner packaging are intact. Do not use masks from a package that is torn, wet or damaged.
  • Take out one mask at a time. Keep the rest in the original carton.
  • Have a lined waste receptacle within reach for removal.
  • Tie back long hair and remove anything that would sit between the mask and your face.

Putting the mask on

1. Wash and dry your hands completely

Use soap and water, or an alcohol-based hand sanitizer. Dry fully before handling the mask — damp hands transfer moisture to the material.

2. Pick the mask up by the ear loops

Hold it by the loops only. Avoid touching the front and inner panels, which are the filtering and face-contact surfaces.

3. Place it on your nose first, then your chin

The stiffer edge with the nose wire goes up, and the pleats open downward. Hook one loop behind each ear, then draw the lower edge down under your chin.

4. Adjust the fit

Pinch the nose wire along the bridge of your nose with both hands until it follows the shape of your face. Run a finger along the edges to close gaps at the cheeks and under the chin. Open the pleats so the mask covers your nose, mouth and chin completely.

5. Avoid touching your face while wearing it

Once the mask is on, leave it alone. If you have to touch or readjust it, clean your hands before and after.

6. Remove the mask using the ear loops

Lean slightly forward, lift both loops off your ears at the same time, and let the mask fall away from your face. Do not touch the front of the mask.

While you are wearing it

  • Keep the mask over your nose, mouth and chin at all times during use.
  • Do not pull it down to your chin or neck and put it back up. A mask moved off the face is contaminated and should be replaced.
  • Do not hang it from one ear, fold it into a pocket, or set it down to put back on later.
  • Replace the mask if it becomes damp, soiled or damaged, if it is difficult to breathe through, or if it is contaminated by splash or spray.
  • In a healthcare setting, change masks between patients and follow your facility’s infection control procedures.

After you take it off

  1. Discard the mask immediately in a waste receptacle. In a healthcare setting, follow your facility’s waste handling procedures — masks used in clinical care may need to be handled as regulated medical waste.
  2. Wash and dry your hands, or use an alcohol-based hand sanitizer, straight after disposal.
  3. Do not wash, launder, disinfect, spray, microwave or otherwise attempt to reprocess the mask. It is a single-use device and is not designed to be cleaned or reused.

Warnings and limitations

  • Single use. Discard after each use. Do not wash, disinfect or reuse.
  • Supplied non-sterile. The mask is not sterile and is not intended for use as a sterile item.
  • Not a respirator. This mask does not provide NIOSH-approved respiratory protection and is not a substitute for an N95 or other filtering facepiece respirator. It is not intended for protection against gases, vapors, chemical fumes, dusts, asbestos, lead or smoke.
  • No mask eliminates risk. Correct use reduces exposure. It does not eliminate the risk of acquiring or transmitting infection or disease, and it should be used alongside hand hygiene and any other infection control practices that apply where you are.
  • Fit limitations. The mask does not form an airtight seal. Facial hair, facial jewelry and anything else between the mask and the skin will reduce how well it sits.
  • One size. The mask is supplied in a single size proportioned for adult faces. No pediatric indication is offered and pediatric fit has not been evaluated. Keep packaging away from infants and small children.
  • Do not use in an MR environment. The mask contains a metal nose wire and must be removed before entering the magnetic resonance imaging suite.
  • Not made with natural rubber latex.
  • Not flame resistant. Do not use near open flame, sparks or high heat.
  • Discontinue use and seek medical advice if you experience skin irritation, difficulty breathing, dizziness or other discomfort while wearing the mask.
  • Do not alter or modify the mask in any way, including cutting, stapling or adding fittings.

Storage and handling

Store masks in the original carton, in a clean, dry indoor space away from direct sunlight, moisture and strong odors. Keep cartons off the floor and away from chemicals, solvents and cleaning agents. Do not store in a vehicle, outdoors, or anywhere subject to temperature extremes or high humidity.

Rotate stock so older cartons are used first. Do not use masks from a carton that has been wet, crushed or contaminated.

Product details

  
Product3-Ply Surgical Mask
ModelUSM1001
ConstructionThree-layer: polypropylene spunbond outer and inner layers, polypropylene meltblown filtration layer
AssemblyUltrasonic welding at the perimeter; no adhesive
RetentionTwo ear loops, polyester yarn with elastane
Nose pieceFormable carbon steel nose wire, polyethylene laminated
Dimensions175 mm × 95 mm (6⅞ in × 3⅝ in) unfolded
ColorWhite
SuppliedSingle-use, non-sterile
LatexNot made with natural rubber latex
Packaging50-count cartons
ManufacturedHouston, Texas, USA

Questions and product complaints

If you have a question about these instructions, or if you need to report a problem with a product you received, contact us with the model number and the lot number printed on the carton:

USA MedPro
2777 Allen Parkway, Suite 920
Houston, TX 77019
713-993-6819 · info@usamedpro.com

Please keep the carton and any remaining masks from the affected lot until we have responded.